Main functions and requirements
The environment condition and operation condition shall be as follows:
Environment temperature: 10℃~ 40℃;
Relative humidity: 30%~75%;
Atmospheric pressure: 700hPa~1060hPa;
- 2
Power source shall be as follows:
Power supply: 220V±22V;
Supply frequency: 50Hz±lHz;
Input power: 35VA;
- 3 Feather:
1) Three channels output
2) Pulse frequency
I (Complete denervation): output pulse frequency 500Hz, modulation wave frequency 0.5Hz-5Hz, adjustable continuously, franchise±15%;
II (Partal denervation): output pulse frequency 0.5Hz-5Hz, adjustable continuously, franchise±15%;
3) Pulse width
I (Complete denervation): pluse width is made up of 5pcs 1ms. Width of modulation wave 10ms, franchise±30%;
II (Partal denervation): pluse width is 10ms;
4) Under 500Ωload resistance, output electricity of each channel ≤80mA, adjustable continuously
5) Treatment time: 5min, 10 min, 15 min, 20 min, 25 min, 30 min, franchise±10%.
6) Output waveform: Two-way asymmetrical square-wave
7) Size: 420*360*232mm, franchise±20mm
Contraindications
1, Amyotrophic lateral sclerosis;
2, The progression of multiple sclerosis progression;
3, The treatment site has a large metal inside body;
4, Malignant tumors;
5, Tuberculosis;
6, Thrombophlebitis;
7, Tetanus